Sensors 2003, 3, 330-339
sensors
ISSN 1424-8220
© 2003 by MDPI
/sensors
Strategies of Miniaturised Reference Electrodes Integrated in a
Silicon Based “one chip” pH Sensor
Anette Simonis 1,2, Hans Lüth 2, Joseph Wang 3, Michael J. Schöning 1,2*
1 University of Applied Sciences Aachen, Ginsterweg 1, 52428 Jülich, Germany
2 Institute of Thin Films and Interfaces, Research Centre Jülich GmbH, 52425 Jülich, Germany
3 New Mexico State University, Las Cruces, NM 88003-800 1, USA
Author to whom correspondence should be addressed.
Received: 07 Ap ril 2003 / Accep ted: 15 July 2003 / Published: 11 Sep tember 2003
Abstract: Different types of Ag/AgCl reference electrodes have been realised by means of
thin- and thick-film technique. For inner electrolyte, KCl-containing membranes have been
deposited and different coatings have been used to protect the reference electrode from a fast
leaching out of KCl. The stability of the potential of the reference electrodes without KCl-
containing membranes in 3 M KCl was about 7 hours for thin-film electrodes and up to 90
hours for thick-film electrodes. The reference electrodes with membrane were stable for
more than two months. The thick-film reference electrode was integrated with a thin-film
pH sensor onto one chip and the system pH sensor/reference electrode was investigated by
means of capacitance/voltage (C/V) and constant capacitance (ConCap) mode.
Keywords: reference electrode, thin-film technique, thick-film technique, “one chip”
sensor, pH sensor.
Introduction