Study on pH Effect in Process of an Entero-gastric Fiber-optic
Sensor Design
1,* 1 2 1
Chen Guo-ping , Xia Rong-min , Gong Jun and Shou Wen-de
1
Department of Biomedical Engineering, Shanghai Jiao Tong University, Shanghai 200030, China.
Tel.86-21 Fax 86-21 E-mail: gpchenapple@
2
Department of Digestion, Second Hospital of Xi’an Jiaotong University, Xi’an 710004,China
* Author to whom correspondence should be addressed.
Received: 28 October 2002 / Accepted: 15 November 2002 / Published: 30 November 2002
Abstract: The pH effect on design of a fiber optic sensor newly developed for monitoring
entero-gastric reflux is investigated. The research has been carried out by utilizing the
sensor to measure characteristic absorption of bilirubin standard solution (used as a
calibrator) at different pH values from about 1 to 8. The results show explicitly that the
estimation error according to the calibration curves with pH 7 is significantly lower than
that at alkaline pH values. Compared with the in vitro experimental data, the optimum pH
range for entero-gastric reflux analysis is proposed to be between 5-6. The possible reasons
for estimation error of the fiber optic sensor are given in detail.
Keywords: Fiber optic sensor, Bilirubin, pH, Entero-gastric reflux