电位滴定法测定氯离子.docx
文本预览下载声明
数据记录和处理原始数据记录和计算NaCl标准溶液称取氯化钠质量 m=0.2941g,容量瓶体积:100ml = = =0.05 mol/L手动标定硝酸银溶液(1)原始数据表格1银离子滴定氯离子数据滴定体积/ml电位/mV滴定体积/ml电位/mV滴定体积/ml电位/mV01722.902184.204430.301763.002214.304480.601803.102274.404500.901823.202344.504531.201863.302394.604561.501903.402464.704581.801853.502575.004632.101993.602775.304672.402053.703545.604702.502063.804005.904732.602113.904236.204752.702134.004336.504772.802144.10438(2)二阶微商法计算滴定终点为方便计算,挑选原始实验数据中的部分数据整理如下:滴定体积/ml电位φ/mVΔV/(-)Δφ/(-)0.30176————1.201860.901011.11—3.102270.904145.5634.453.202340.1077024.443.302390.10550-203.402460502570.1011110403.602770.1020200903.703540.10777705703.804000.1046460-3103.904230.1023230-2304.004330.1010100-1304.104380.10550-505.304671.202924.17-25.836.504771.202016.67-7.5去掉一个坏值(3.30,239),找出二阶微商为零的两个数据点(3.70,354)和(3.80,400)。代入公式= +(—)* 有,计量点电位为:= 354+ 46* = 383.80mV自动电位滴定标定硝酸银浓度起始体积:0.01mL;终点体积:3.49mL。终点电位:392mV消耗硝酸银体积:V1=3.49—0.01=3.48mL50mL自来水水样测定起始体积:3.50mL;终点体积:3.62mL消耗硝酸银体积:V2=3.62—3.50=0.12mL数据处理硝酸银浓度计算由硝酸银和氯化钠的反应关系式有:3*0.05=3.48*C故,硝酸银的浓度C==0.043mol/L自来水中氯离子的测定由 0.043*0.12=50*C,有自来水氯离子的含量为:C= =1.02*mol/L思考题理论计算计量点电位:384mV,实际滴定电位:392mV。二者差异原因:当接近计量点时,电位变化越来越大。可能前一滴溶液未到计量点电位,仪器将自动再继续缓慢滴加溶液,此时1滴(约为0.02mL)硝酸银就可能导致体系电位超过所设定的滴定终点。做好本实验的建议1、准确记录实验数据;2、准确计算计量点电位。07105005
显示全部