标准操作说明书手册-DSC-v11.pdf
TheLifeSciencesandBio-Engineering
CoreFacility
OperatingInstructions–DifferentialScanningCalorimeter
ModelandManufacturer:DSC6000,PerkinElmer
Safety:
-Weargloveswhenoperatingtheinstrument.
-InformtheCoreFacilitystaffforanysafetyissueormisuseoftheinstrument.
Note:
1.Itisbesttoknowyoursamplepropertiesinadvance.Themaximumtemperatureofthetestshallnot
exceedthedecompositiontemperatureofthesample(toavoidirreparablelosscausedbythepollutionof
decompositionvolatilestothetestingfurnace).
1.Samplepreparation
1.1Weighasampleof3-6mg.
1.2Usetweezerstoputthesampleintothepan.
1.3Coverthepanwithitscoverandputontotheblackholderofthepressinstrument.
1.4Rotatetheleverofthesamplepressinstrumenttopressyoursample.(seepicturebelow)
PanandCoverBlackHolderwithPanSamplePressInstrument
2.HardwareStart-up
2.1Turnonthecomputer.
2.2Openthemainnitrogengasvalve(blue)inananti-clockwisedirection.Adjusttheoutputvalve(red
circle)tokeepthepressurewithintherange0.15-0.25MPa.
NitrogenGasCylinder0.2MpaonGasOutputMeter
2.3TurnontheDSCinstrumentbyswitchingonthebuttonbehindtheinstrument.Thepowerlight
shouldturnon.
InstrumentMainSwitchPowerLightisOn
TheLifeSciencesandBio-Engineering