贝朗Dialog+血液透析机透析液回路结构与功能分析.pdf
文本预览下载声明
原理与应用 Instrument Theory ·123·
贝朗Dialog+ 血液透析机透析液回路结构与功能分析
高 爽
摘要 论述了贝朗 +血液透析机透析液回路,包括透析液除气加热、透析液配比、超滤以及消毒等过程的构成
[ ] Dialog
与功能,分析了有关功能对透析患者的影响,以提高操作者或护士的工作效率,达到为患者实施安全高效血液透析的
目的。
关键词 血液透析机;透析液回路;透析液;超滤;消毒
[ ]
中国图书资料分类号 文献标识码 文章编号 ( )
[ ] TH772. 1 [ ] B [ ] 1003-8868 2011 06-0123-04
Analysis of Structures and Functions of B.Braun Dialog+ Dialysate Circuit
GAO Shuang
(The 309 Hospital of PLA , Beijing 100091, China)
Abstract In this article the constructions and functions of the dialysate circuit of the Dialog + machine, consisted of
heating, degassing, dialysate proportioning , ultrafiltration, and disinfection systems, are analyzed and discussed
comprehensively, and particularly the effects of the functions and parameters of the circuit on the patients are emphasized,
with the aim to enhance the performance of the operators or nurses in terms of the hemodialysis safety and clinical
, , ( ): ,
outcomes of the patients.[Chinese Medical Equipment Journal 2011 32 6 123-125 131]
Key words hemodialysis machine; dialysate circuit; dialysate; ultrafiltration; disinfection
引言 高、中、低 挡。当进入水达到 高水位时,进水电磁阀关
1 3 NSVD
血液透析机透析液回路系统的主要功能是配制成治疗 闭;处于NSVD 中水位(低于高位触点)时,打开进水阀;而降
所需的透析液;调节透析液流量和压力;维持透析液适当的 至 低水位触点时,即自动停止加热棒 加热。 水箱
NSVD
显示全部