Blood flow ml/min 50 CVVH Adult Aqualine Recirculation mode 0:00 h.min -250 200 Access 46 mmHg -80 300 Return TMP 63 mmHg -150 200 24 mmHg 11 mmHg -50 450 Pr.Drop Go to connection Go to programming More Help End treatment Remove tube system -250 200 Blood flow ml/min 50 CVVH Adult Aqualine Start connection Access 52 mmHg -80 300 Return TMP 57 mmHg -150 200 25 mmHg 6 mmHg -50 450 Pr.Drop Single connection Double connection Go to recirculation Help Go to programming -250 200 Blood flow ml/min 50 CVVH Adult Aqualine Confirm window Access 52 mmHg -80 300 Return TMP 57 mmHg -150 200 25 mmHg 6 mmHg -50 450 YOU HAVE SELECTED SINGLE CONNECTION MODE By selecting this connection mode, you have to connect the access line only, fill the circuit with blood until blood detected alarm appears and then connect return line and start treatment. With this connection mode, the patient will lose volume(priming volume of the circuit+filter). A blood flow rate of 50 ml/min has to be used to fill the circuit with blood and then increased in incremental steps after start treatment selection. Select yes-to use this single connection mode No-to go back to the previous screen Pr.Drop No Yes -250 200 Blood flow ml/min 50 CVVH Adult Aqualine Confirm window Access 52 mmHg -80