fluke-1663-multifunction-tester-使用操作说明书手册-supplement.pdf
ManualSupplement
ManualTitle:1662/1663/1664FCUsersSupplementIssue:3
PrintDate:October2015IssueDate:11/17
Revision/Date:1,4/16PageCount:2
Thissupplementcontainsinformationnecessarytoensuretheaccuracyofthe
abovemanual.
©2016-2017FlukeCorporation.Allrightsreserved.x
�
1662/1663/1664FCUsersManualSupplement
Change#1,579
Onpage74,underGeneralSpecifications,addthefollowingundertheElectromagnetic
Compatibility(EMC)
SomemobiledevicesthattransmitRFenergymaytransmitlevelsthatfarexceed3V/mandmaydamagesensitiveelectronic
circuits.Toinsurethebestperformance,donotallowadevicewhichistransmittingRFenergyinexcessof3V/mtobewithin
30cmoftheTesterwhileinuse.
Change#2,594
Onpage84,replacetheTestSignalstablewith:
TestSignals
RCDTypeTestSignalDescription
ACThewaveformisasinewavestartingatzerocrossing,polarity
(sinusoidal)determinedbyphaseselection(0°phasestartswithlowtohighzero
crossing,180°phasestartswithhightolowzerocrossing).The
magnitudeofthetestcurrentisInxMultiplierforalltests.
∆
AThewaveformisahalfwaverectifiedsinewavestartingatzero,
(halfwave)polaritydeterminedbyphaseselection(0°phasestartswithlowto
highzerocrossing,180°phasestartswithhighto