BA 119F/00/en/06.03 silometer Software Version 1.x FMX 570 Level Measurement Operating Instructions FMX 570 Endress+ Hauser The Power of Know How Silometer FMX 570 Measurement at a Glance Level Measurement at a Glance Function Matrix Action 1 Reset transmitter V9H5 Enter 671: »+« and »-« keys, ⇒ changes digit Press »E« to register entry - Omit if commissioned as in Section 4.1 2 »Empty« calibration* V0H1 Fill vessel 0…40% full (probe covered) Enter level in %, m, ft, etc. Press »E« to register entry 3 »Full« calibration* V0H2 Fill vessel 60…100% full Enter level in %, m, ft, etc. Press »E« to register entry 4 0/4 mA signal V0H3 Enter 0 for 0…20 mA signal, 1 for 4…20 mA signal Press »E« to register entry V0H5 Enter level for 0/4 mA signal (if not 0) Press »E« to register entry V0H6 Enter level for 20 mA signal (if not 100)