前臂多腱肌的多通道表面肌电检测电极阵列装置的设计_英文_杨丹丹.pdf
文本预览下载声明
第 10 卷 第 2 期 纳 米 技 术 与 精 密 工 程 Vol. 10 No. 2
2012 年 3 月 杨丹丹等: 前臂多腱肌的多通道表面肌 电检测电极阵列装置的设计( 英文) · ·
2012 年 3 月 Nanotechnology and Precision Engineering Mar. 2012
Design of Electrode Array for Multichannel sEMG Recording of
Multitendoned Forearm Muscles
YANG Dan-dan ,HOU Wen-sheng ,WU Xiao-ying ,ZHENG Xiao-lin
( Bioengineering College ,Chongqing University ,Chongqing 400044 ,China)
Abstract: A linear electrode array was designed to record multichannel surface electromyography
( sEMG) of multitendoned forearm muscles noninvasively. Eight gold-coated bar electrodes were fabrica-
ted using a flex-print technique and arranged in one-dimensional array with inter-electrode distance( IED)
( center-to-center) of 5 mm. The disposable adhesive non-contact structure was made with pre-punched
double-sided medical tape for filling a small amount of conductive cream and attaching to skin with ease.
Single differential sEMG signal was amplified and band-pass filtered. Low electrode impedance guaran-
tees the consistent electrical property of gold-coated electrode. Low and relatively stable electrode-skin
impedances ensure the stability of baseline and a good signal-to-noise ratio( SNR) . To explore the practi-
cal possibility of the electrode array ,14 channel sEMG signals of nine subjects were simultaneously recor-
ded from flexor digitorum superficialis( FDS) during single-digit force production tasks. High stability of
baseline ,excellent quality of sEMG signals and good channel discrimination of muscle activation indicate
that this electrode array device is a promising technology for long-